Biophysical Characterisation

Bio-Layer Interferometry (BLI)

A label-free, high-throughput analytical technique for real-time monitoring of biomolecular interactions. It provides information on binding kinetics and affinity and is particularly well suited for antigen–antibody interaction studies.

Application
Characterization of protein-protein and protein-ligand interactions, including binding affinity and association/dissociation rates. Particularly suitable for high-throughput antibody screening, epitope binning and cross-competition assays to identify antibodies targeting the same or overlapping epitopes.

Equipment: Octet R8, Sartorius

Description
Bio-Layer Interferometry measures changes in the interference pattern of light reflected from a biosensor surface. One of the interacting molecules (ligand) is immobilized on the biosensor surface, while the other molecule (analyte) is introduced in solution. Different types of biosensors and immobilization strategies are available, depending on the nature of the molecule and the experimental objective. Biomolecular binding at the sensor surface generates a signal proportional to the change in molecular layer thickness.

Sample requirements
Approximately 200 µL per sample well. The required analyte concentration depends on the interaction being studied and the expected affinity.